TLQ

TLQ FAMILY Single phase transformer for operating rooms' spotlights.

TLQ

Dry type transformer

IP-20 enclosure in last generation V-0 fireproof polymer box, according to UL94

Safety class I, convertible in class II

Includes LED functioning indicator and electrostatic screen

Full power capacity in every conection 

Optional protection fuse

Possibility of tailor-made manufacturing

Showing page to
 

TLQ160

TLQ315

TLQ630

TLQ800

Rating160 VA315 VA630 VA800 VA
Input voltage230 V±5%230 V±5%230 V±5%230 V±5%
Output voltage24-25-26-27 V24-25-26-27 V24-25-26-27 V24-25-26-27 V
Frequency50/60 Hz50/60 Hz50/60 Hz50/60 Hz
Protection degreeIP-20IP-20IP-20IP-20
CoolingANANANANANANANAN
Ambient temperature25ºC25ºC25ºC25ºC
Temperature riseClass BClass BClass BClass B
InsulationClass B - 130ºCClass B - 130ºCClass B - 130ºCClass B - 130ºC
WindingsClass HC - 220ºCClass HC - 220ºCClass HC - 220ºCClass HC - 220ºC
Test voltage2,5 kV (1 min, 50 Hz)2,5 kV (1 min, 50 Hz)2,5 kV (1 min, 50 Hz)2,5 kV (1 min, 50 Hz)
StandardsIEC/EN/UNE-EN 61558-1, CEIEC/EN/UNE-EN 61558-1, CEIEC/EN/UNE-EN 61558-1, CEIEC/EN/UNE-EN 61558-1, CE
MountingScrews and DIN RailScrewsScrewsScrews
Weight2,3 kg4,1 kg6,8 kg10 kg
Dimensions106x123x118x110x74 mm 5Ø118x138x131x121x88 mm 6Ø136x162x156x145x104 mm 6Ø136x162x180x145x104 mm 6Ø
  ADDADDADDADD